Symbols, display messages and troubleshooting
141
Symbols, display messages and troubleshooting
9
This message/symbol
appears…
and means:
The test strip is stuck.
The test strip drum does not rotate.
Press the Test button to turn the
meter off. Open the drum compart-
ment and, if necessary, pull out
the test strip. Press the red drum
release button and remove the test
strip drum. Insert the test strip drum
and close the drum compartment.
The meter then rotates the drum
and reads the bar code. The drum
symbol and bottle symbol then flash
in the display. The 90-day use-by
period will no longer be properly
calculated for this drum. Perform a
quality control check. You may use
up all the test strips remaining in the
drum. The drum symbol continues
to flash in the display until you have
inserted a new drum.
This message/symbol
appears…
and means:
The meter has been exposed to
severe differences in temperature.
The temperature is too low (between
0 °C – +6 °C) or too high (above
+44 °C) to perform a test.
Press the Test button to turn the
meter off. Let it cool down or warm
up slowly. During testing the per-
mitted temperature range is from
+10 °C – +40 °C.
You applied too little blood or control
solution.
Turn the meter off and on again.
Repeat the test with a new test strip
and a larger amount of blood or con-
trol solution.
There is no test strip drum in the
meter.
Press the Test button to turn the
meter off. Insert a new test strip
drum.
