110
Using the meter as an electronic notebook
111
Using the meter as an electronic notebook
5
When you release the button, the display reverts to the average.
30-day period
14-day period
7-day period
Memory location 1
Most recent result saved
7-day period
Lowest value Average Highest value
Lowest value Average Highest value
Lowest value Average Highest value
You can exit from the average value display at any time:
To do this, press the Test button.
The meter turns off.
Notes
i
Results flagged as performance checks with the bottle symbol
are not included in the calculations.
i
Results shown as being or are not included in the average
calculations, but are displayed as the lowest and highest values,
respectively.
i
The meter calculates the average value, the lowest value and the
highest value for a period even when the saved results cover a
shorter period. Example: tests were only performed in the last 5
days. In addition, either no other results were saved or the results
that were saved are more than 30 days old. Consequently the
results for the last 5 days are used for calculating the average,
the lowest value and the highest value for all three periods (7, 14
and 30 days).
+
+
+
+