Chapter 4 44
Performing a Control Test
You need the meter, a test strip, and control solution Level 1 or Level 2.
The control level is printed on the bottle label.
1. Insert a test strip into the
meter in the direction of
the arrow. The meter
turns on.
2. Make sure the code number
on the display matches the
code number on the test
strip container. If you miss
seeing the code number,
take the test strip out and
reinsert it into the meter.
3. Select the control solution
you want to test. You will
enter the level later in the
test.