Resumo do Conteúdo

Página 1 - ACCU-CU-CHEKHEK

BLOOD GLUCOSE METEROwner’s BookletACCU-CU-CHEKHEK®Aviva

Página 2

Chapter 1: Understanding Your New System 8▲!Coding Your MeterA new code key is inside each new box of test strips.1. Make sure your meter is off. 2. T

Página 3 - System Owner

Chapter 1: Understanding Your New System9Notes:• Don’t force the code key into the meter — it is made to go in the meter only one way.• If you see a c

Página 4 - Aviva System

Chapter 1: Understanding Your New System 101.Press and release to turnthe meter on. The time and dateappear on the display. “Set” andthe hour flash.

Página 5 - Health-Related Information

Chapter 1: Understanding Your New System114. Repeat steps 2 and 3 to setthe minutes, am/pm, month,day, and year. The flashingfield is the one you arec

Página 6 - Contents

Chapter 1: Understanding Your New System 12Using ACCU-CHEK Aviva Test Strips• Only use ACCU-CHEK Aviva test strips.• Replace the code key and run a co

Página 7

Chapter 1: Your New Meter 13Chapter 2: Control TestingWhy Run Control TestsRunning a control test lets you know that your meter and test strips are wo

Página 8 - The ACCU-CHEK Aviva Meter

Chapter 2: Control Testing 14Running a Control TestYou need the meter, a test strip, and control solution Level 1 or Level 2.The control level is prin

Página 9 - Chapter 1: Your New Meter 7

Chapter 2: Control Testing152. Make sure the code number onthe display matches the codenumber on the test stripcontainer. If you miss seeing thecode n

Página 10 - Coding Your Meter

Chapter 2: Control Testing 16||||||||||||Running a Control Test (continued)6. Squeeze the bottle until a tiny drop forms at thetip of the bottle. Touc

Página 11

Chapter 2: Control Testing178. Press to set the level in the meter.9. The control result and “OK” alternate on thedisplay if the result is in range.

Página 13

Chapter 2: Control Testing 18Understanding Control Test ResultsExampleThe label on your test strip container shows the acceptableranges for both the L

Página 14

Chapter 2: Control Testing 19If your control result is not inside the acceptable range (printed on your test strip container), here are somethings you

Página 15 - Chapter 2: Control Testing

Chapter 3: Testing Your Blood Sugar 20Chapter 3: Testing Your Blood SugarUsing the ACCU-CHEK Multiclix Lancet DeviceNotes:• Testing with blood from pl

Página 16 - Running a Control Test

Chapter 3: Testing Your Blood Sugar21Inserting a Lancet DrumYou must first load the drum into the lancet device to get it ready for use.1. Remove the

Página 17

Chapter 3: Testing Your Blood Sugar 223. Slide the cap on until it stops by aligning thenotch on the cap with the notch on the device.4. You are now r

Página 18

Chapter 3: Testing Your Blood Sugar23Preparing the Lancet Device for Fingertip Testing▲!Use a new lancet for each fingerstick to avoid infection and g

Página 19

Chapter 3: Testing Your Blood Sugar243. Press the plunger in as far as it will go, like a pen. 4. The release button turns yellow. Do not press therel

Página 20 - Level 2 255–345

Chapter 3: Testing Your Blood Sugar 25Running a Blood Sugar Test with Blood from Your FingertipBefore you run your first blood test, set up your meter

Página 21 - Troubleshooting Checks

Running a Blood Sugar Test with Blood from Your Fingertip (continued)Chapter 3: Testing Your Blood Sugar265. Hold the lancet device firmlyagainst the

Página 22

Chapter 3: Testing Your Blood Sugar 27Note: Do not twist the plunger when the release button is yellow or when the lancet counter showsno white bars.

Página 23 - Inserting a Lancet Drum

Dear ACCU-CHEK®System Owner,Thank you for choosing the ACCU-CHEK®Aviva system!Congratulations on your decision to take control of your diabetes. We’ve

Página 24

Chapter 3: Testing Your Blood Sugar 28Changing the Lancet DrumWhen you have used the sixth and last lancet, you should change the drum.1. Take the blu

Página 25

Chapter 3: Testing Your Blood Sugar29Note: A used drum cannot be reinserted into the lancet device.In rare instances, if a used drum is dropped on a

Página 26

IMPORTANT• Talk to your doctor before you begin using other test sites.Chapter 3: Testing Your Blood Sugar 30Running a Blood Sugar Test with Blood fro

Página 27

Chapter 3: Testing Your Blood Sugar 31▲!Non-fingertip testing may be done:• Immediately before a meal• FastingFingertip test only:• Two hours or less

Página 28 - Yellow release

Running a Blood Sugar Test (non-fingertip) (continued)You need the meter, a test strip, a lancet device with a drum loaded, and the clear cap in your

Página 29

Chapter 3: Testing Your Blood Sugar 333. Dial the depth setting to 5 1/2 for forearm, upperarm, thigh, or calf testing.Dial the depth setting to 3for

Página 30 - Changing the Lancet Drum

Running a Blood Sugar Test (non-fingertip) (continued)Chapter 3: Testing Your Blood Sugar346. Make sure the code number on the displaymatches the code

Página 31

Chapter 3: Testing Your Blood Sugar35Yellow releasebutton8. Press the opening of the clear cap firmlyagainst a fleshy area on the non-fingertip site.P

Página 32 - IMPORTANT

Chapter 3: Testing Your Blood Sugar 3610. Touch the drop to the tip of the yellow windowof the test strip. Do not put blood on top ofthe test strip.11

Página 33

Chapter 3: Testing Your Blood Sugar3712. Your result appears on the display. If you want to flag your test result for a special event, leavethe test s

Página 34

The ACCU-CHEK®Aviva SystemYour new ACCU-CHEK Aviva meter and accessories work together to measure the amount of sugar(glucose) in your blood to help y

Página 35

14. The lancet counter decreases by onewhite bar. For safety reasons, once youadvance to a new lancet, you cannot goback to a used lancet.Chapter 3: T

Página 36

Chapter 1: Your New Meter 39Flagging Test ResultsYou might want to flag an event such as exercise or a non-fingertip result. If you wish, you can “fla

Página 37

Chapter 3: Testing Your Blood Sugar40Understanding Your Test ResultsThe ACCU-CHEK Aviva test strips are whole-blood referenced and have been calibrate

Página 38

Chapter 3: Testing Your Blood Sugar 41Symptoms of High or Low Blood SugarBeing aware of the symptoms of high or low blood sugar can help you understan

Página 39

Chapter 3: Testing Your Blood Sugar 42Comparing Your Meter Result to a Lab ResultA common question is how the blood sugar results on your meter compar

Página 40 - Having trouble testing

Chapter 4: Meter Memory, Setup, and Downloading43Notes:• Do not change your therapy based on one individual result in memory.• The memory is not lost

Página 41 - Flagging Test Results

Chapter 4: Meter Memory, Setup, and Downloading44Viewing Test Results1. Press either or to enter memory. Your most recent result appears.2. Press t

Página 42 - Unusual Test Results

Chapter 4: Meter Memory, Setup, and Downloading45Meter SetupUsing the Set ModeBy using the set mode, you can personalize your meter to suit your lifes

Página 43

Chapter 4: Meter Memory, Setup, and Downloading46Setting the Time and DateTIME / DATE(hrs, min, am/pm,month, day, year)BEEPER(on/off)TEST REMINDERS(A-

Página 44

Chapter 4: Meter Memory, Setup, and Downloading475. Repeat steps 3 and 4 to set the minutes,am/pm, month, day, and year. The flashing fieldis the one

Página 45 - Storing Test Results

Important Information About Your New Meter• Your new meter is designed and approved for testing fresh capillary whole blood samples (for example,blood

Página 46 - Viewing Test Results

Chapter 4: Meter Memory, Setup, and Downloading48Setting the Beeper On/OffYour new meter has the beeper preset to “On.” You can set the beeperto “OFF,

Página 47 - Meter Setup

Chapter 4: Meter Memory, Setup, and Downloading49|||||||||||||||||||||||2. To enter the setmode, press and holdfor about fourseconds. “Set” flasheson

Página 48 - Setting the Time and Date

Chapter 4: Meter Memory, Setup, and Downloading50Setting Test RemindersTest reminders are a handy way to remind you to test. You can set 1 to 4 remind

Página 49

Chapter 4: Meter Memory, Setup, and Downloading511. Press to turn the meter on.The flashing test strip symbolappears.2. To enter the set mode, pressa

Página 50 - Setting the Beeper On/Off

Chapter 4: Meter Memory, Setup, and Downloading52Setting Test Reminders (continued)4. Press and release or to select “On” or “OFF.” Pressand release

Página 51

Chapter 4: Meter Memory, Setup, and Downloading538. Press and release to setminutes. Am or pm flashes onthe display.9. Press or to switchbetween am a

Página 52 - Setting Test Reminders

Chapter 4: Meter Memory, Setup, and Downloading54Setting the Hypoglycemic (Hypo) IndicatorYou can set the meter to let you know when your result indic

Página 53

Chapter 4: Meter Memory, Setup, and Downloading551. Press to turn the meteron. The flashing test stripsymbol appears.2. To enter the set mode, pressa

Página 54

Chapter 4: Meter Memory, Setup, and Downloading564. Press and release or to switchbetween “On” and“OFF.” Press andrelease to setyour choice.5. If you

Página 55

Chapter 4: Meter Memory, Setup, and Downloading57Downloading Your Results to a ComputerWe offer a variety of software to help you download your result

Página 56

ContentsChapter 1: Understanding Your New System ...6The ACCU-CHEK Aviva Meter

Página 57

Chapter 4: Meter Memory, Setup, and Downloading585. Point the two IR windows toward each other. They should be 1 to 4 inches (3-10 cm) apart.6. With t

Página 58

Chapter 1: Your New Meter 59Chapter 5: Maintenance and Troubleshooting59Chapter 5: Maintenance and TroubleshootingChanging the Battery1. Open the batt

Página 59 - Transferring Data

Chapter 5: Maintenance and Troubleshooting60Cleaning Your MeterCaring for your ACCU-CHEK Aviva meter is easy - just keep it free of dust. If you need

Página 60

Chapter 1: Your New Meter 61Chapter 5: Maintenance and Troubleshooting61Your meter needs little or no maintenance with normal use. Itautomatically tes

Página 61 - Changing the Battery

Chapter 5: Maintenance and Troubleshooting 62▲!Screen Messages and TroubleshootingNever make treatment decisions based on an error message. If you hav

Página 62 - Cleaning Your Meter

Chapter 5: Maintenance and Troubleshooting63The meter is ready for you toinsert a test strip.The meter is ready for a drop ofblood or control solution

Página 63 - Maintenance and Testing

Chapter 5: Maintenance and Troubleshooting64The meter isn’t coded or thecode key isn’t all the way in.Turn off the meter and recode it.See Chapter 1 “

Página 64

Chapter 5: Maintenance and Troubleshooting65The test strips will expire at the endof the current month. Before the endof the month, insert a new code

Página 65

Chapter 6: Technical Information 66Chapter 6: Technical InformationProduct LimitationsPlease read the literature packaged with your test strips to fin

Página 66

Chapter 6: Technical Information67Relative humidity operating range 10 to 90%Memory capacity 500 results with time and dateDimensions 94 x 53 x 22 mm

Página 67

Meter Setup...45Setting the Time

Página 68 - Specifications

Chapter 6: Technical Information68ACCU-CHEK Aviva Limited 3-Year WarrantyRoche Diagnostics warrants to the original purchaser of the meter that your A

Página 69 - Product Safety Information

Chapter 6: Technical Information69Some states do not allow limitations on how long an implied warranty will last or the exclusion of incidentalor cons

Página 70

Chapter 6: Technical Information70▲!CAUTION—A RESTRICTED LICENSE LIMITS USE OF THE ACCU-CHEK®AVIVA SYSTEM—READ CAREFULLY THE LIMITATIONS RECITED BELOW

Página 71 - Non-Warranty Service Policy

CAUTION—A RESTRICTED LICENSE LIMITS USE OF THE ACCU-CHEK®MULTICLIX DEVICE(lancet device and lancet drum).The ACCU-CHEK Multiclix device and its use ar

Página 72 - AVIVA SYSTEM

Chapter 6: Technical Information72Additional SuppliesThe following supplies and accessories are available from your authorized Roche Diagnostics diabe

Página 73

Chapter 6: Technical Information73Information for Healthcare ProfessionalsHealthcare Professionals: Follow the infection control procedures appropriat

Página 74 - Additional Supplies

Chapter 6: Technical Information 74References1. Stedman,Thomas Lathrop.Stedman’s Medical Dictionary, 27th Edition,1999, pg. 2082.2. American Diabetes

Página 75

Indexappetite, increased, 41battery, changing, 59battery, type, 59beeper settings, 48blood sugar, testing, 20blurred vision, 41button, on/off, 6code k

Página 76 - References

troubleshooting, 62 – 65use by, 12, 13warranty, 68 – 69Index 76

Página 78 - Index 76

Chapter 1: Understanding Your New System*The ACCU-CHEK Aviva MeterDisplay –Shows results,messages, andresults stored inmemory.Right and LeftArrow Butt

Página 79

ACCU-CHEK®Aviva©2005 Roche Diagnostics. All rights reserved. 04447344003-0605Assembled and distributed in the USA by Roche Diagnostics, Indianapolis,

Página 80 - ACCU-CHEK

Chapter 1: Your New Meter 7Test Strip Container Top ViewOn/Off/SetButtonInfrared (IR) WindowYellowWindow –Touch blooddrop or controlsolution here.Comf

Comentários a estes Manuais