Symbols, error messages and troubleshooting
Farbe: P 021 – Prüfmittelnummer: 201
This error message ... means:

O After testing, you inserted a new test
strip while End was still being dis-
played.
Remove the test strip by pushing the
ejector down and re-insert it. After
testing, wait for the meter to turn off
before inserting a new test strip.
If this error message appears frequently or continuously, the meter is
defective. Please contact your local customer support and service
centre (see Chapter 19).
O The test strip drew up too little blood
or control solution.
Push the ejector down to eject the
test strip. Repeat the test with a new
test strip and a larger amount of
blood or control solution.
