17
Chapter 2: Control Testing
Why Perform a Control Test
Performing a control test lets you know that the meter and test strips are working properly. You should
perform a control test when:
You open a new box of test strips
You left the test strip container open or you think the test strips have been damaged
The test strips were stored in extreme temperatures and/or humidity
You want to check the meter and test strips
You dropped the meter
Your test result does not match how you feel
You want to check if you are testing correctly