Press f to confirm the bolus amount and
duration.
You hear a melody and the bolus delivery begins
within the next 3 minutes.
Note During the entire bolus delivery, the remaining time and
amount of the Extended bolus and the current hourly basal
rate are displayed in the RUN screen.
Note If a temporary basal rate is active at the same time, the
remaining time and amount of the bolus and the amount of
the increased or decreased hourly basal rate are displayed in
the RUN screen.
8.1.2 Cancel an Extended bolus
During programming (the bolus amount or bolus duration blinks):
You have three options for cancelling an Extended bolus during
programming:
If you do not press any key for 20 seconds h, your insulin pump
returns to the RUN screen (timeout).
Exit to the EXTENDED BOLUS menu by pressing d+a
simultaneously.
Set the bolus amount to 0.0 units and press f.
No bolus is delivered.