Accu-Chek Spirit Guia do Utilizador Página 67

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 66
Prepare the pump for use
65
FC_prime_EN.eps
start
priming
start-up
procedure
stop
priming
STOP
screen
prime the
infusion set
3 sec.
Make sure your insulin pump is in STOP.
Press d to move to the PRIME THE INFUSION
SET menu.
Select with f.
START PRIMING and the priming amount of
25 units h of insulin is shown on the display.
Press and hold f until you hear a melody and
the piston rod moves forward. The priming
process starts and a count-up is shown on the
display. Your insulin pump automatically stops
priming after 25 units h of insulin.
When the priming process is finished, your
insulin pump returns to STOP.
The infusion set is properly primed when no air
bubbles are visible in the tubing anymore and a
bubble-free flow of insulin emerges from the tip
of the needle.
Vista de página 66
1 2 ... 62 63 64 65 66 67 68 69 70 71 72 ... 225 226

Comentários a estes Manuais

Sem comentários