Accu-Chek Spirit Guia do Utilizador Página 199

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 198
Technical data
197
* Note: these factory settings may vary in some countries and depend on the changes you,
your doctor or healthcare team made using your insulin pump or pump programming
software for healthcare professionals.
Basal rate
hourly basal rate + max.
TBR (hourly basal rate
combined with temporary
basal rate, max. increase)
hourly basal rates
increment
– hourly basal rates
temporary basal rate
TBR increase or decrease
in increments of
TBR duration
TBR duration increments
– basal rate profiles available
– basal rate lock
Bolus
max. bolus amount per bolus
delivery
bolus amount increments
for “Quick” Standard bolus
bolus duration (Extended
bolus, MultiWave bolus)
in increments of
Prime volume
Maximum parameter range
adjustable with the Accu-Chek
Insulin Pump Configuration
Software
max. amount
0 – 62.50 IU/h
max. amount
0 – 25.0 IU/h
0 – 250%
10%
15 min – 24 h
15 min, 30 min, 1 h
1 – 5
on or off
max. amount
0 – 25.0 IU
0.1, 0.2, 0.5, 1.0 or 2.0 IU
15 min – 24 h
15 min, 30 min, 1 h
max. amount
0 – 30.0 IU
Typical range available
on the pump and
standard parameter
settings*
20.0 IU/h
0.1 IU
10.0 IU/h
0 – 200%
10%
15 min – 24 h
15 min
1 – 5
on or off
presetting: off
25.0 IU
0.1, 0.2, 0.5, 1.0 or 2.0 IU
presetting: 0.5 IU
15 min – 12 h
15 min
0 – 30.0 IU
presetting: 25 IU