Technical data
193
Note 1 At 80 MHz and 800 MHz, the higher frequency range applies.
Note 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from
structures, objects and people.
a Field strengths from fixed transmitters, such as base stations for
radio (cellular/cordless) telephones and land mobile radios, amateur
radio, AM an FM radio broadcast and TV broadcast cannot be
predicted theoretically with accuracy. To asses the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the
location in which the Accu-Chek Spirit is used exceeds the applicable
RF compliance level above, the Accu-Chek Spirit should be observed
to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or
relocating the Accu-Chek Spirit.
b Over the frequency range 150 kHz to 80 MHz, field strengths
should be less than 10 V/m.
V
1
in V:
Rated maximum
output power
of transmitter W
10.00
150 kHz to 80 MHz
out of ISM
(Industrial, Scientific,
Medical Band)
E
1
in V/m:
80 MHz to
800 MHz
10.00
800 MHz to
2.5 GHz
0.01
0.1
1
10
100
0.04
0.11
0.35
1.11
3.50
0.04
0.11
0.35
1.11
3.50
0.07
0.22
0.70
2.21
7.00
Separation distance according to frequency of transmitter m